Does the MANTECH Buret conform to Class A buret specifications?

Yes! Details available here for MANTECH burets.

Yes, MANTECH burets conform to Class A specifications as per ASTM E287 - Standard Specification for Laboratory Glass Graduated Burets. The standard defines the requirements for Class A and B burets as follows:

TABLE 1 Requirements for Burets

    Graduated Scale    
Capacity,
mL
Subdivision,
mL
Class A
and B
Class A
Scale Length,
mm
0 to Capacity
Class B
Scale Length,
mm
0 to Capacity
Distance, Top
of Buret to "0"
Graduation,
mm

A

Class A
and B
Number
Each, mL
Class A
Volumetric
Tolerance,
±mL
Class B
Volumetric
Tolerance,
±mL
 10  0.05 350 to 450 250 to 450 30 to 75 0.5 0.02 0.04
 25 0.1 350 to 450 250 to 450 30 to 75 1 0.03 0.06
 50 0.1 500 to 600 500 to 600 40 to 100 1 0.05 0.10
100 0.2 550 to 650 550 to 650 40 to 100 2 0.10 0.20

MANTECH burets are available with 10mL and 25mL syringes, and have a resolution of roughly 100,000 steps providing precise injections down to the microliter level. Actual performance of burets and syringes are verified in MANTECH's QC process with accepted tolerances exceeding requirements of ASTM E287, and a report is provided with each new syringe. 

Users can perform their own verification of dispense volume accuracy by navigating to Hardware>Manual Control>Buret in MANTECH Pro software.