Due to manual entry of sample name, volume, and potentially DO/Temperature information (if entering this information manually as well), there is a risk of typos/incorrect information entered into the run. Therefore, being able to change and update this information after the run has completed is critical.
Procedure:
- Navigate to Template → BOD Method.
- Load BOD5, or the active BOD Method used in your BOD Run sequence.
- Click on Settings (Gear Icon).
- There will be two sections under the Incubation header for:
- Max re-read time (hours)
- Max Sample Editing Time (hours)
- Extend the time period to the desired values, and click Save to ensure the settings on this menu is saved.
- Then click Save Method.