Why do I experience communication issues with IntelliDose™ (metering pump) after using prime button on pump?
Typical error to appear is 'Communication error! Cannot synchronize computer and drive baud rates'.
Think of the pump have two modes, where:
- When powering the pump ON the first time, the pump is in Auto mode.
- When the Prime button on the pump is used (or pressed by accident), it switches the pump into Manual mode.
- Even though turning off the prime button stops the pump, the mode is still set to Manual.
Therefore, power-cycle the pump to regain automation control.